3.2 Basic Measurement Example
25
3
Chapte
r 3
Measur
em
en
t
Measurement
12
Connect the test leads to a battery.
Open-terminal voltages for the instrument are as follows:
3 m
and 30 m
ranges: 25 V peak
300 m
range: 7 V peak
3
to 3000
: 4 V peak
These voltages derive from the load associated with charging the 1.2 uF
capacitor inside the instrument.
• The open-terminal voltage for the 3 m
, 30 m
, and 300 m
ranges peaks at
4 V approximately 500 ms after the terminal is placed in the open state.
• When building a measurement line using scanners, use a relay with a dielec-
tric strength that is greater than or equal to the open-terminal voltage for the
range being used.
13
Read the measured resistance and voltage.
See
"3.7 Displaying Measurement Results" (
See
Measured resistance
value
Measured voltage
value
Please refer to "Measured value is unstable." of "Before returning for repair." (
p.173) as a measurement and attention.
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...