8.7 Basic Data Importing Methods
157
8
Chapt
er 8
RS-232
C
/GP-I
B
In
te
rf
aces
Flexible data importing is available depending on the application.
8.7 Basic Data Importing Methods
Free-Run Data Importing
Initial
Setup
:INITiate:CONTinuous ON
(enable continuous
measurement)
:TRIGger:SOURce IMM
(internal triggering)
Importing
:FETCh?
Imports the most recent measurement value
Importing by Host Triggering
Initial
Setup
:INITiate:CONTinuous OFF
(disable continuous
measurement)
:TRIGger:SOURce IMM
(internal triggering)
Importing
:READ?
A trigger occurs, and a measurement is taken and the result is trans-
ferred.
Importing Data by TRIG Key or
TRIG
Terminal
Initial
Setup
:INITiate:CONTinuous OFF
(disable continuous
measurement)
:TRIGger:SOURce EXT
(external triggering)
Importing
:READ?
When triggered by the
TRIG
key or TRIG terminal, a measurement is
taken and the result is transferred.
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...