4.1 Comparator Function
51
4
Chapt
er
4
Appl
ie
d Measur
ement
Two methods are available for setting comparator thresholds.
Comparator Threshold Method Selection
1
(The SHIFT indicator lights up.)
The Comparator setting display appears.
2
Press so that the indicated position blinks, and set the comparator threshold
method.
HIGH, LOW
..... Compare against specified upper and lower thresholds (default setting
method)
REF, %
............ Compare against upper and lower thresholds internally calculated from a
specified reference value and tolerance
About comparisons based on a reference value and tolerance
When the reference value and tolerance method is selected, thresholds are calculated as
follows:
Upper threshold = reference value
×
(100 + tolerance [%]) / 100
Lower threshold = reference value
×
(100 - tolerance [%]) / 100
Measured values are displayed as a percentage relative to the reference value, calcu-
lated as follows:
Relative value = (measured value - reference value) / reference value
×
100 [%]
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...