8.6 Message Reference
150
Execute a Measurement and Read the Measured Values
Syntax
Query
:READ?
Response
<Measured resistance (NR3)>, <Measured voltage (NR3)>
(
V mode)
<Measured resistance (NR3)>
(
mode)
<Measured voltage (NR3)>
(V mode)
Description
Query
Switches from the Idle State to the Trigger Wait State, then reads the
next measured value. With auto-ranging enabled, the most suitable
range is selected before measurement.
Example
Query
:READ?
Response
289.68E-3, 1.3921E+0
(
V mode)
Measured resistance is 289.68 m
, and voltage is 1.3921 V.
Error
This command causes an execution error if issued during the Continuous Mea-
surement state (after
:INITIATE:CONTINUOUS ON
).
Note
• The next command does not execute until measurement is finished.
• When the trigger source is external, the
TRG
command does not trigger mea-
surement.
• Wait 100 ms before applying the trigger with
:READ?
immediately after
changing the measuring conditions during measurement.
See
" Measurement Value Formats" (
Trigger Source Operation
IMMediate
Triggers and reads measured value.
EXTernal
After triggering by the TRIG terminal (EXT I/O) or
TRIG
key,
reads the measured value.
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...