The ERR output signal indicates the occurrence of measurement fault conditions
(such as open test leads, or a bad contact).
There are two ERR output methods.
Synchronized
with EOM Out-
put (SYNC)
Measurement faults detected while measuring (not while awaiting trigger or
during delay or calculation intervals), are indicated by ERR output synchronous
with EOM output (the end-of-measurement signal).
ERR Output Low (On):
A measurement fault has prevented correct measure-
ment
ERR Output High (Off): Correct measurement obtained (OF or -OF: Out-of-
range cases are included)
Asynchronous
with EOM Out-
put (ASYNC)
Measurement faults (test lead connection conditions) are output in real time. The
output is asynchronous with the TRIG signal and EOM output.
ERR Output Low (On):
Measurement fault condition (open test leads, or a bad
contact)
ERR Output High (Off): Test lead connections are normal
ERR Output
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...