2.5 Selecting the Line Frequency
20
The instrument's power supply frequency must be set in order to eliminate noise.
Although the power supply frequency setting is configured automatically ("AUTO") by default, it can also be
set manually. Measured values will not be stable if the power supply frequency is not set properly.
2.5 Selecting the Line Frequency
1
(The SHIFT indicator lights up.)
The Menu display appears.
2
Select the line frequency setting display.
See
"1.4 Menu Display Sequence (SHIFT > ENTER)" (
3
Select the frequency of the AC mains supply being used.
AUTO
.... Automatic configuration of power supply frequency
50
.......... 50 Hz
60
.......... 60 Hz
4
Applies the settings and returns to the measurement display.
(Main display)
(Sub display)
(Main display)
(Sub display)
(Main display)
(Sub display) flashing
• When set to automatic configuration (AUTO), a power supply frequency of
either 50 Hz or 60 Hz will be automatically detected whenever the instru-
ment is turned on or reset.
• Changes in the power supply frequency occurring at other times will not be
detected.
• The power supply frequency will be set to either 50 Hz or 60 Hz, whichever
is closer.
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...