Specifications
into service according to the EMC information supplied in the tables that
follow.
E
LECTROMAGNETIC
I
MMUNITY
G
UIDANCE
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The P006979 is intended for use in the electromagnetic environment specified
below. The customer or the user of the model P006979 should make sure it is
used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment—
Guidance
RF emissions
CISPR 11
Group 1
The P006979 is an intentional radiator
of RF energy. Therefore, its RF emissions
are low and may cause interference in
nearby electronic equipment (see the
tables that follow).
RF Emissions
CISPR 11
Class A
NOTE: The EMISSIONS characteristics of
this equipment make it suitable for use
in industrial areas and hospitals (CISPR
11 class A). If it is used in a residential
environment (for which CISPR 11 class B
is normally required) this equipment
might not offer adequate protection to
radio-frequency communication ser-
vices. The user might need to take miti-
gation measures, such as relocating or
re-orienting the equipment
.
Harmonic Emis-
sions
IEC 61000-3-2
Not applica-
ble
Voltage Fluctua-
tions/
Flicker Emissions
IEC 61000-3-3
Not applica-
ble
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P006979 is intended for use in the electromagnetic environment specified
below. The customer or the user of the P006979 should make sure it is used in
such an environment.
Immunity Test
IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic
Environment—
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 2 kV, ± 4 kV, ±
8 kV, and ± 15
kV Air
± 8 kV Contact
± 2 kV, ± 4 kV, ±
8 kV, and ± 15
kV Air
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material,
the relative humid-
ity should be at least
30%.