WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
30
1.2 Power Cable Replacement
1.2
Power Cable Replacement
1.
Make sure the bed is unplugged and you have done the steps for
“Setup” on page 25.
2.
Get access to the Logic Control P.C. board (VersaCare® Bed, A through
J model bed) or upper control board (UCB) (VersaCare® Bed, K or
newer model bed or Progressa® Bed). Refer to the bed’s service
manual.
3.
Make a note of the power cable connections on the
Logic Control
P.C. board
or UCB, and then disconnect the power cable from the
Logic Control P.C. board
or UCB.
4.
Connect the
new
power cable to the connectors on the
Logic
Control P.C. board
or UCB.
•
Progressa® Bed and VersaCare® Bed, K or newer model
—the
power cable connects to P14.
•
VersaCare® Bed, A through J model
—the power cable
connects to P1 and P11.
5.
Make a note of the power cable routing from the
Logic Control P.C.
board
or UCB to the reader, and then use the same routing for the
new
power cable. Remove and install cable ties as necessary.
NOTES:
•
VersaCare® Bed
—you will
need to remove the cable
channel cover for access to
the cable routing in the
channel.
•
Progressa® Bed
—you will
need to remove the hose
cover screw and hose cover
for access to the cable
routing along the air hose.
6.
Near the reader, remove the cable tie(s) that holds the power cable
with the other cables on the bed.
DRAFT 26-FEB-2019
Summary of Contents for WatchCare Series