1.2 Power Cable Replacement
13. Install the P.C. board in the
reader.
WARNING:
Warning
—Failure to correctly align the gaskets on the covers of the
reader could permit fluids to get into the enclosure. Injury or equipment
damage could occur.
14. Install the back cover on to the
reader.
Make sure the gasket
inserts are correctly aligned.
15. Install the six screws to attach the back cover to the reader.
16. Remove any tape residue from the back cover of the reader.
17. Use the alcohol-based cleaner to clean the back cover of the reader
and the area where the reader was installed on the bed. Let the back
cover and area dry.
18. Remove one side of the adhesive covering from the double-sided
tape gasket, and install the gasket on to the back cover.
19. Remove the adhesive covering
from the other side of the
gasket, and install the reader on
the bed so that its label is
toward the outside of the bed.
20. Firmly press the reader into position for 10 seconds with
approximately 75 lb (34 kg) of force.
21. At the cable tie mount(s) near the reader, install a cable tie(s) to hold
the cables on to the bed.
DRAFT 26-FEB-2019