1.2 Power Cable Replacement
22. Do as applicable for your bed:
VersaCare® Bed
a.
Install the cable channel
cover.
b.
Install the cover for the
Logic Control P.C. board
(VersaCare® Bed A through
J model) or the UCB
(VersaCare® Bed K or newer
model). Refer to the bed’s
service manual.
Progressa® Bed
a.
Put the hose cover in
position on the bed, and
install the screw to attach
the hose cover to the bed.
b.
Install the cover for the
UCB. Refer to the bed’s
service manual.
23. The replacement is complete. Do the function checks on page 53
and in the bed’s service manual. As you operate the bed’s articulation
functions, make sure the cables are not in a location where they
could get damaged or pinched.