WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
50
1.6 Equipment Call Cable Replacement
9.
Remove the six screws that
attach the back cover to the
reader.
CAUTION:
Caution
—Failure to wear an antistatic strap could cause equipment
damage.
10. Put on the antistatic strap.
11. Remove the P.C. board from the reader.
12. Disconnect the
old
equipment
call cable from J17 on the P.C.
board, and remove the cable
from the bed.
13. Connect the
new
equipment call cable to J17 on the P.C. board.
14. Install the P.C. board in the
reader.
DRAFT 26-FEB-2019