WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
43
1.4 Antenna Assembly Replacement
CAUTION:
Caution
—Make sure to install the new antenna in the correct orientation
and location. Otherwise, the antenna may not operate as intended.
5.
Remove the adhesive covering from the gasket on the bottom of the
new
antenna, and then install the antenna in the correct orientation
and location on the sleep deck. Firmly press the antenna into
position for 10 seconds with approximately 75 lb (34 kg) of force.
6.
Follow the routing of the antenna cable to the reader, and then use
the same routing for the
new
antenna cable. Remove and install
cable ties as necessary.
7.
Progressa® Bed
—for the head section antenna, put a service loop in
the cable, and
loosely
secure the new cable tie to allow cable
movement (see below).
With StayInPlace™ Feature
Without StayInPlace™ Feature
NOTE:
To determine if your bed has the StayInPlace™ Feature, look at the bottom
of the Progressa® Therapy label on the intermediate siderail. A bed with
the StayInPlace™ Feature will have the StayInPlace™ logo on the label,
8.
Near the reader, remove the cable tie(s) that holds the antenna cable
with the other cables on the bed.
9.
At the antenna-reader cable to
antenna cable connection,
remove the heat shrink or
insulated tape.
10. Disconnect the antenna-reader cable from the antenna cable.