3.7.3
Change the Nominal Values in stand still or during the run
Note:
- Multiple parameters can be modified by the explained procedure.
1. Lock user-input to LOCK 5: SELECT to 00633 (set Lowbyte, Bit 7)
PC EOT
ADR
STX
0 0 6 3 3 = 0 0 8 0
ETX
BCC
Cent ADR
ACK
PC EOT
2. Write new nominal parameter / parameters:
006XX = 00601, 00603, 00606, 00611, 00612, 00617, 00618 or 00620
PC EOT
ADR
STX
0 0 6 x x = V V V V
ETX
BCC
Cent ADR
ACK
PC EOT
3. Execute the change of the nominal value / values (and keep LOCK 5 set).
PC EOT
ADR
STX
0 0 6 3 3 = 0 0 8 8
ETX
BCC
Cent ADR
ACK
PC EOT
FYI: Is the nominal value out of the input-range, in SIOF (00685) bit 7 is set and the centrifuge answers with:
Cent ADR
NAK
Then ENQUIRY of parameter 00685 is necessary
PC EOT
A 0 0 6 8 5
ENQ
Cent ADR
STX
0 0 6 8 5 = V V V V
ETX
BCC
PC EOT
The SIOF-Error bit is erased after enquire the SIOF and other parameters can be changed by SELECT.
After the change of one / or of several parameters, the centrifuge indicates the actual program by marking
the program number "in brackets".
Example: Program-No. before the change: 1
Program-No. after the change: ( 1)