EN
53/70
3.7.1
Sequence after starting PC communcation
1 Serial data
Description
E :EOT:T:00685:ENQ
:T:STX:00685=0000:ETX:[05]
Read SIOF , activate "serial commands"
Clear reset bit etc …
E :EOT:T:00537:ENQ
:T:STX:00537=C800:ETX:[07]
Read the centrifuge ID , to get the centrifuge type
Highbyte: Centrifuge type
Lowbyte: Type of cooling
Centrifuge type C800 = Rotanta 460 + pos
i.E. Centrifuge type F800 = SK 20.08
E :EOT:T:00528:ENQ
:T:STX:00528=1800:ETX:[08]
Read the positioning and hatch state ,
Highbyte: = Hatch state , hatch is closed , lid lock of hatch is closed
Lowbyte = Positioning state
E :EOT:T:00634:ENQ
:T:STX:00634=0162:ETX:[0A]
Read Centrifuge state 1
Highbyte: Just called program no is 1
Lowbyte : Centrifuge state = stand still
(Bit 5 and 6 only for internal use )
E :EOT:T:00635:ENQ
:T:STX:00635=0292:ETX:[07]
Read Centrifuge state 2
Highbyte : Not relevant ,
Lowbyte : Rotor number 9, 2 = Key-Lock 2
E :EOT:T:00524:ENQ
:T:STX:00524=0602:ETX:[0A]
Read target rotor position
Highbyte : Maximum number of rotorpositions = 6,
Lowbyte : Number of actual rotorpositions = 2