2.10.1.4 Miscellaneous parameters part 1
Parameter-No.: 00512
Show RCF or RPM mode in the display
Gen. 2
Access:
RW
Structure:
MSD LSD MSD LSD
Lowbyte: RCF / RPM mode
Highbyte: Not relevant
Lowbyte : RCF / RPM mode
7
6
5
4
3
2
1
not relevant
not relevant
not relevant
not relevant
not relevant
not relevant
not relevant
0
1 = rcf mode 0 = rpm mode
If the radius (parameter 00620) is changed, the display will automatically show the RCF mode.
If the set speed (parameter 00603) is changed, the display will automatically show the RPM mode.
Parameter-No.: 00513
Dual Timing Mode
Gen. 2
Access:
R
Structure:
MSD LSD MSD LSD
Lowbyte: Dual timing mode on/off
Highbyte: Not relevant
Lowbyte : Dual timing mode
7
6
5
4
3
2
1
not relevant
not relevant
not relevant
not relevant
not relevant
not relevant
not relevant
0
1 = dual timing mode on 0 = dual timing mode off
Dual timing mode
on
means : Running time counter starts when set speed is reached.
Dual timing mode
off
means : Running ting time counter starts at start of centrifugation