EN
35/70
Parameter-No.: 00618
Set-Temperature
Gen. 1 + 2
Access:
RW
Structure:
MSD LSD MSD LSD
Lowbyte: Set temperature
Highbyte: Always 00
Input:
Temperature in degree Celsius
Range:
Cooled centrifuges:
-20 up to +40 °C
Heated/Cooled centrifuges: -20 up to +60 °C
Format:
Data is transmitted standardized: with offset 25° (T
min
= 25)
See also 00619 actual temperature
Formula:
Transmitted data = (Set value + Tmin) *2
Parameter-No.: 00619
Temperature-actual value
Gen. 1 + 2
Access:
R
Structure:
MSD LSD MSD LSD
Lowbyte: Actual temperature
Highbyte: Always 00
Display:
Temperature value in Celsius
Range:
-25 up to +102 °C
Format:
Data is transmitted standardised: with offset 25° (T
min
= 25)
see also 00619 actual temperature
Formula:
transmitted data = (Set value - T
min
) *2
e.g.: -25°C
=
0X00
-10°C
=
0X1E
+40°C
=
0X82
+102°C
=
0XFE
Parameter-No.: 00620
Radius
Gen. 1 + 2
Access:
RW
Structure:
MSD LSD MSD LSD
Lowbyte: Radius (LSB)
Highbyte: Radius (MSB)
Range:
rotor radius [mm] from 10 up to 330.
The range of the radius is not checked by the centrifuge. The correct radius has to be limited by the PC:
After setting the radius, display is always in RCF mode