![Hettich ROTANTA 46 RSC ROBOTIC Operating Instructions Manual Download Page 24](http://html1.mh-extra.com/html/hettich/rotanta-46-rsc-robotic/rotanta-46-rsc-robotic_operating-instructions-manual_2127795024.webp)
EN
24/70
Parameter-No.: 00523
Recall or store a program
Gen. 2
Access:
W
Requirement: Key-Lock is in position LOCK 2
Structure:
MSD LSD MSD LSD
Lowbyte: Recall or store command
Highbyte: Program-number (0 up to 89)
Highbyte: program-number (0 up to 89 RCL , 1 up to 89 STO)
7
1 =
(not relevant)
6 MSb
5
4
3
2
1
0 LSb
7-bit digit, which indicates the program-number to recall or store
Lowbyte: program command
7
6
5
4
3
2
1
0
7-bit digit, which indicates the program command
Lowbyte = 01 : RCL
Stored
program
from
the
EEPROM will be copied to the edit block
Lowbyte = 04 : RCL & activate
Stored program from the EEPROM will be copied to the active/current program block
and make this program block active
Lowbyte = 08 : STO
Copy the edit block to the EEPROM
Lowbyte = 18h : STO & activate
Copy the edit block to the EEPROM and to the active/current program block
and make this program block active.
Explanation see chapter 3.6 "How to store and recall programs".
Note: a program is a "set of parameters".