Parameter-No.: 00633
Control-command
Gen. 1 + 2
Access:
RW
Structure:
MSD LSD MSD LSD
Lowbyte:
Control-command
Highbyte: Command for Report
Highbyte:
7 1 =
Report requested
6 1 =
Report terminated
5 1 =
not relevant
4 1 =
not relevant
3 1 =
not relevant
2 1 =
not relevant
1 1 =
not relevant
0 1 =
not relevant
Bit "Report desired" has a static function, i.e. it is set by the PC, it has to be waited after
each
centrifugation for bit
"terminated". Then the centrifuge can be restarted again.
Lowbyte: Control-Command
7 1 =
LOCK 5 (Input locked, excluding STOP)
6 1 =
LOCK 4 (Input locked = LOCK 2)
5 1 =
not relevant
4 1 =
not relevant
3 1 =
Modification of nominal value is required
2 1 =
not relevant
1 1 =
START centrifugation
0 1 =
STOP centrifugation