![Hettich ROTANTA 46 RSC ROBOTIC Operating Instructions Manual Download Page 29](http://html1.mh-extra.com/html/hettich/rotanta-46-rsc-robotic/rotanta-46-rsc-robotic_operating-instructions-manual_2127795029.webp)
EN
29/70
2.10.2 Parameters for Generation 1 and Generation 2
Parameter-No.: 00601
Set run-time
Gen. 1 + 2
Access:
RW
Structure:
MSD LSD MSD LSD
time
LSB
time
MSB
Input: Seconds
Range:
0 = continuous run, time run = 1 up to 59,999 sec (999 min : 59 sec)
Parameter-No.: 00602
Actual run-time
Gen. 1 + 2
Access:
R
Structure:
MSD LSD MSD LSD
time
LSB
time
MSB
Display: Seconds
Range:
0 up to 59,999 sec (999 min:59 sec)
Parameter-No.: 00603
Set-speed
Gen. 1 + 2
Access:
RW
Structure:
MSD LSD MSD LSD
Set
speed
LSB
Set speed MSB
Input:
RPM = revolutions per minute
Range:
50 up to the maximum rotor speed (see parameter 00605)
Parameter-No.: 00604
Actual-speed
Gen. 1 + 2
Access:
R
Structure:
MSD LSD MSD LSD
Actual speed LSB
Actual
speed
MSB
Display: RPM
=
R
evolutions
P
er
M
inute
Range:
50 up to 20,000 RPM