33
16. TROUBLESHOOTING
ERROR CODE
DESCRIPTION
CORRECTIVE ACTION
Er 7
The lid was
opened during
test/check
Press button and remove the
cartridge. Insert a new cartridge and
perform a new test.
Er 14
(Do not re-use)
An already used
Test cartridge or
Monthly Check
cartridge is used
(may occur when
re-using Monthly
Check cartridge)
Press button and remove the
cartridge. Insert a new cartridge and
perform a new test.
Lo°C
The ambient
temperature is
too low
Ensure the equipment is within normal
operating temperatures for at least 10
minutes. Insert a new cartridge and
perform a new test.
Hi°C
The ambient
temperature is
too high
Ensure the equipment is within normal
operating temperatures for at least 10
minutes. Insert a new cartridge and
perform a new test.
<4%
The HbA1c test
result is lower
than 4.0%
(20mmol/mol)
Use the Daily Check cartridge to confirm
the analyzer performance. Insert a new
cartridge and perform a new test.
>14%
The HbA1c test
result is higher
than 14.0%
(130mmol/mol)
Use the Daily Check cartridge to confirm
the analyzer performance. Insert a new
cartridge and perform a new test.
The
Analyzer
does not
start
There is a
problem with the
power adaptor or
the analyzer itself
Unplug the power adaptor and
reconnect. If the problem still persists,
contact your local distributor.
GB
Summary of Contents for HbA1c 501
Page 1: ...Operating Manual HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 2: ......
Page 43: ...41 MEMO GB ...
Page 45: ...Bedienungsanleitung HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 46: ......
Page 87: ...41 MEMO DE ...
Page 89: ...Gebruiksaanwijzing HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 90: ......
Page 131: ...41 MEMO NL ...
Page 133: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 134: ......
Page 175: ...41 MEMO SE ...
Page 177: ...Brugsanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 178: ......
Page 219: ...41 NOTATER DK ...
Page 221: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 222: ......
Page 263: ...41 NOTATER NO ...