13
11. RUN a HbA1c TEST
Reagent pack sampling area
Use of Venous Blood
Venous whole blood collected in tubes with K2EDTA, lithium heparin, sodium
citrate or sodium fluoride/oxalate as an anticoagulant can be used.
Venous whole blood can be stored at 2-8°C for 7 seven days with unbroken seal (only
3 days when seal is broken) and at 20-25°C for 3 days.
Mix the tube thoroughly prior to testing. Open the lid and take out one drop of
blood with a pipette. Place a drop of blood on a hydrophobic surface. Gently touch
the drop of blood with the tip of the sampling area of the reagent pack. Ensure that
the sampling area is completely filled.
NOTE:
Do not wipe off excess blood outside the sampling area.
Do not touch the open end of the sampling area.
CAUTION:
There is a potential risk of biological hazard.
All parts of the HemoCue® HbA1c 501 System should be considered
potentially infectious.
•
Use gloves
•
Dispose used test cartridges in a sturdy container with lid
•
Consult local environmental authorities for proper disinfection
procedures as well as disposal of consumables
[adequate]
[inadequate]
sampling area
GB
Summary of Contents for HbA1c 501
Page 1: ...Operating Manual HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 2: ......
Page 43: ...41 MEMO GB ...
Page 45: ...Bedienungsanleitung HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 46: ......
Page 87: ...41 MEMO DE ...
Page 89: ...Gebruiksaanwijzing HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 90: ......
Page 131: ...41 MEMO NL ...
Page 133: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 134: ......
Page 175: ...41 MEMO SE ...
Page 177: ...Brugsanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 178: ......
Page 219: ...41 NOTATER DK ...
Page 221: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 222: ......
Page 263: ...41 NOTATER NO ...