18. MAINTENANCE
Cleaning the analyzer
1) Disconnect the analyzer before cleaning.
2) Wipe off the Analyzer using a soft, clean cloth.
3) Do not use liquid or aerosol cleaners.
4) If necessary, an alcohol swab can be used to remove dirt, other chemical
solutions should not be used. Cleaning solutions may cause damage to the
analyzer surface.
Cleaning the cartridge compartment of the analyzer
STEP 1.
Turn the power off and disconnect the power cord before cleaning.
STEP 2.
When the lid is opened,
press the two black knobs to rotate
the cartridge compartment forward.
STEP 3.
Moisten a sponge with ethanol or recommended commercial disinfectant
(Recommended disinfectants: CaviCide®, METREX® RESEARCH COPORATION, US)
STEP 4.
Gently wipe off the cartridge compartment using the sponge.
Do not to drip any liquids into the system.
Replacing the Fan Filter:
Regularly check the fan filters located on the back of the analyzer. Replace
the fan filter if clogged by dust. Extra filters are included at delivery of the analyzer.
NOTE:
Additional fan filters can be purchased from your local distributor.
STEP 1.
Remove
the fan filter cover.
STEP 2.
Take out
the used fan filter -
replace with a new.
STEP 3.
Close the
cover.
Filter
Filter cover
GB
Summary of Contents for HbA1c 501
Page 1: ...Operating Manual HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 2: ......
Page 43: ...41 MEMO GB ...
Page 45: ...Bedienungsanleitung HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 46: ......
Page 87: ...41 MEMO DE ...
Page 89: ...Gebruiksaanwijzing HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 90: ......
Page 131: ...41 MEMO NL ...
Page 133: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 134: ......
Page 175: ...41 MEMO SE ...
Page 177: ...Brugsanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 178: ......
Page 219: ...41 NOTATER DK ...
Page 221: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 222: ......
Page 263: ...41 NOTATER NO ...