20
13. SYSTEM CHECK
Monthly Check Cartridge Monthly Check Reagent Pack
When to Use the Monthly Check Cartridge
•
Once a month before samples are tested.
•
On suspicion that test results may be incorrect.
•
After an error message (Er 4 or Er 5).
The Monthly Check Cartridge, used to check the Analyzer, consists of a cartridge
and a regent pack, same as the test cartridge, but without the sampling area.
CAUTION:
•
Do not reuse.
•
Do not use Monthly Check Cartridge after expiration date.
NOTE:
•
If Monthly Check Cartridges are reused, the analyzer will display ´Er 14
(Do not reuse)´
•
The Monthly Check Cartridge can be purchased from your local
distributor.
How to use the Monthly Check Cartridge
1) Open the lid of the Analyzer.
Rinsing Solution
Reagent Solution
Safety Guard