24
13. SYSTEM CHECK
3) Mix the reagent pack gently 5-6 times and touch the control solution with
the tip of the sampling area.The control solution is automatically drawn up into
the sampling area. Wait until the sampling area is completely filled with control
solution.
CAUTION:
Do not mix the reagent pack too vigorously, it may cause air
bubbles. If bubbles are present, wait until they disappear.
IMPORTANT:
Analysis must begin immediately after the reagent pack is filled
with control solution.
4) Insert the reagent pack into the cartridge compartment of the analyzer.
NOTE:
Do not
contaminate the
sampling area of the
reagent pack.
5) Close the lid. The test starts automatically