12
11. RUN a HbA1c TEST
STEP 5-1.
Gently mix by turning the reagent pack back and forth 5 - 6 times before applying
blood sample.
CAUTION:
Do not mix too vigorously, it may cause air bubbles. If bubbles are
present wait until they disappear.
Capillary whole blood from fingertip and venous whole blood can be used for
HbA1c test. 4 μL capillary whole blood from fingertip or venous whole blood is
needed for HbA1c testing
STEP 5-2.
Apply the blood sample by gently touching the drop of blood with the tip of the
sampling area. Ensure that the sampling area is completely filled.
IMPORTANT:
Place the reagent pack in the cartridge compartment and start
test within 30 seconds once blood has been applied.
Sample Collection and Handling
- Use of Capillary Blood
Puncture the fingertip to get a minimum of 4μL of capillary blood sample. Gently
touch the blood sample with the tip of the sampling area of the Reagent Pack. The
blood is automatically drawn in via capillary action.
Ensure that the sampling area is completely filled.
Summary of Contents for HbA1c 501
Page 1: ...Operating Manual HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 2: ......
Page 43: ...41 MEMO GB ...
Page 45: ...Bedienungsanleitung HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 46: ......
Page 87: ...41 MEMO DE ...
Page 89: ...Gebruiksaanwijzing HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 90: ......
Page 131: ...41 MEMO NL ...
Page 133: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 134: ......
Page 175: ...41 MEMO SE ...
Page 177: ...Brugsanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 178: ......
Page 219: ...41 NOTATER DK ...
Page 221: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 222: ......
Page 263: ...41 NOTATER NO ...