23
13. SYSTEM CHECK
8) Remove cartridge when the test is completed.
NOTE:
•
If an error message is displayed, repeat the test.
•
Do not use the analyzer if an error message persists.
•
Contact your local distributor for support.
•
To leave Monthly Check Cartridge mode, press .
Quality Control - External control material
The system can be verified according to local guidelines using a commercially
available HbA1c control.
If required, the verification is recommended in connection with installation or when
training a new operator.
Recommended control solutions are, level 1 and level 2 of Lyphochek® Diabetes
Control from Bio-rad or CueSee® HbA1c from Eurotrol.
Only human whole blood controls can be used with the HemoCue® HbA1c 501
system. Contact your local distributor for technical support.
How to test with the Control Solution
1) Open the lid of the analyzer.
2) Insert the cartridge when ‘Insert Cartridge’ is displayed.
GB
Summary of Contents for HbA1c 501
Page 1: ...Operating Manual HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 2: ......
Page 43: ...41 MEMO GB ...
Page 45: ...Bedienungsanleitung HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 46: ......
Page 87: ...41 MEMO DE ...
Page 89: ...Gebruiksaanwijzing HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 90: ......
Page 131: ...41 MEMO NL ...
Page 133: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 134: ......
Page 175: ...41 MEMO SE ...
Page 177: ...Brugsanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 178: ......
Page 219: ...41 NOTATER DK ...
Page 221: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Page 222: ......
Page 263: ...41 NOTATER NO ...