Installation & Configuration
8 .5
Factory Default Settings
Factory settings may be simultaneously returned to factory default values.
NOTE
►
The Factory Default Settings may not be the same as the settings that were factory-calibrated before the Compartmental
Access Refrigerator was shipped.
►
If Factory Default Settings are restored, all calibration values and User PINs will be lost.
General Settings
Restored Value
Home Screen Application Icons
i.C³ APPS, Settings, Temperature Graph, Automatic Alarm Test,
Information Logs
Temperature Setpoint
4.0 ºC
Password (for Settings screen)
1234
Password Protection (for Settings screen)
On
Sounds
On
Alarm Volume
9
Alarm Tone
2
Unit ID
No default value
Date Format
MM/DD/YY
Day
Not affected (maintained in real-time clock)
Month
Year
Time Format
12-hour
Minute
Not affected (maintained in real-time clock)
Hour
AM/PM
Language
English
Temperature Units
ºC
Temperature Graph Screensaver
Off
Alternate Screensaver
On
Min/Max Temperature Display
On
Air Temperature Display
On
Air Temperature Alarm
On
Display Brightness
High (3 symbols)
Temperature Calibration Settings
Values entered at the factory
Access Control Touchpad
On