Installation & Configuration
8 .3 Settings
Current settings may be viewed and changed through the i.C³ monitoring and control system. The i.C³
temperature monitor and controller is programmed at the factory.
Changing temperature settings affects operation of the HaemoBank™. Do not change settings unless instructed in product
documentation or by Haemonetics
Corporation BloodTrack
Customer Support.
NOTE
►
If the Settings screen is password protected or if viewing settings for the first time, enter factory default password of “1234”.
►
Default values for general settings, alarm settings, and display settings are available in Appendix B of the i.C³ User Guide.
Settings screens
To change a setting first access the Settings screen, then the Setting.
8 .3 .1
Temperature Settings
Temperature settings are programmed at the factory. The Temperature Setpoint can be accessed, viewed and
changed through the i.C³ Settings screen by touching the Settings icon and entering the current password. If
accessing the i.C
3
for the first time, use the factory-originated password (1234).
8 .3 .1 .1 Temperature Setpoint
NOTE
►
Default chamber temperature setpoint is 4.0 ºC.
►
Setpoint can be changed from the i.C
3
and from the Haemonetics
®
BloodTrack Courier
®
.
The temperature setpoint is the desired temperature of the HaemoBank™ for stored product.
Change the temperature setpoint if:
►
Your organization requires a chamber temperature other than 4.0 ºC
Perform the following:
1
Touch
Home
,
i.C³ APPS
,
Settings
.
2
Enter the Settings password.
3
Touch
+
or
–
on the
Temperature Setpoint
spin box to select desired setting.
4
Touch
Home
.