360203-1/B
28
Maintenance & Service
4
Lower the front edge of the tray cover.
5
With one hand on each side of the tray, near the cover, gently spread the top edges of the tray outward.
6
While holding the top edges of the tray outward, use your thumbs to press the forward edge of the tray cover
downward.
►
The tabs on the outer edges of the tray cover will engage the corresponding slots on the top edges of the
tray
Tab on outer edge engaged with slot
7
Push the tray completely into the compartment location until it locks.
8
Close the exterior door.
Refer to the BloodTrack
Courier
®
User Guide for details on returning the HaemoBank™ refrigerator to normal use.
10 .3 .2 Replace Tray Bumpers
NOTE
►
Only use tray bumpers supplied with the Compartmental Access Refrigerator. The use of non-Compartmental Access
Refrigerator tray bumpers may allow the tray to be removed without following the proper protocols of the BloodTrack
®
system.
►
Contact Haemonetics
®
Corporation BloodTrack
®
Customer support for replacement bumpers. Part number 800267-1.
Remove a tray bumper:
1
Remove the tray from the compartment location.
(Refer to Section III, Item 10 .3 .1)
.
2
Slide the rubber tray bumper toward the side of the Compartmental Access Refrigerator.
3
Lift the tray bumper upward to disengage it from the keyhole in the bottom of the compartment location.
4
Discard the tray bumper.
Install a tray bumper:
1
Insert the post on the new rubber tray bumper through the wide end of the keyhole in the bottom of the
compartment location.
2
Slide the tray bumper to the narrow end of the keyhole.
3
Grip the tray bumper and pull it upward to ensure it is installed correctly.
►
The tray bumper should not be able to be removed
►
If the tray bumper can be removed by pulling it upward, ensure it is toward the small hole of the keyhole
4
Reinstall the tray in the compartment location.