360203-1/B
21
Maintenance & Service
9.1.3 Power Failure Alarm Test
1
Change the Power Failure delay setting to 0 minutes.
a
Touch
Home
,
i.C³ APPS
,
Settings
.
b
Enter the Settings password.
c
Touch
Alarm Settings
.
d
Touch
–
on the Power Failure spin box to change the value to 0.
2
Switch the AC ON/OFF switch
OFF
.
►
Power Failure alarm will activate immediately
3
Switch the AC ON/OFF switch
ON
.
►
Power Failure alarm will clear and audible alarm will cease
4
Change the Power Failure time delay to the original setting.
a
Touch
Home
,
i.C³ APPS
,
Settings
.
b
Enter the Settings password.
c
Touch
Alarm Settings
.
d
Touch
+
on the Power Failure spin box to change the value to the original setting.
5
Touch
Home
.
9 .1 .4
Door Open Alarm Test
1
Change the Door Open (Time) delay setting to 0 minutes.
a
Touch
Home
,
i.C³ APPS
,
Settings
.
b
Enter the Settings password.
c
Touch
Alarm Settings
.
d
Touch
–
on the Door Open (Time) spin box to change the value to 0.
2
Open the door.
►
Door Open alarm will activate immediately
3
Close the door.
►
Door Open alarm will clear and audible alarm will cease
4
Change the Door Open (Time) setting to the original setting.
a
Touch
Home
,
i.C³ APPS
,
Settings
.
b
Enter the Settings password.
c
Touch
Alarm Settings
.
d
Touch
+
on the Door Open (Time) spin box to change the value to the original setting.
5
Touch
Home
.
9.2
Upgrade System Firmware
Helmer may occasionally issue updates for the i.C³ firmware. Follow the upgrade instructions included with the
firmware update.
9 .3
Test and Replace i .C
3
Monitoring System / Access Control Backup Battery
Test the i .C
3
Monitoring System / Access Control backup battery:
1
Switch the AC ON/OFF switch
OFF
.
►
i.C³ screen should continue to display information with reduced brightness
►
The battery icon will appear on the screen
►
If the Low Battery alarm activates, or the display is blank, replace the battery
2
Switch the AC ON/OFF switch
ON
.