360203-1/B
22
Maintenance & Service
9 .3 .1
Replace the i .C
3
Monitoring System / Access Control Backup Battery
Replacement battery must meet the following specifications: 12V, 7Ah, rechargeable, sealed, lead acid. Contact Haemonetics
®
Corporation BloodTrack
®
Customer Support for replacement battery. Service Kit Part number 800251-1.
Required tools:
►
#2 Phillips screwdriver
Replace the battery:
1
Switch the i.C
3
Monitoring System / Access Control backup battery ON/OFF switch
OFF
. Switch the AC ON/
OFF switch
OFF
. Disconnect the AC power cord from the power receptacle.
2
Using a #2 Phillips screwdriver, remove 6 screws securing the front base cover.
3
Disconnect the power wires from the battery terminals.
4
Using a #2 Phillips screwdriver, remove the screws securing the battery strap to the base of the unit.
5
Remove the strap and set it aside.
6
Remove the battery from the HaemoBank™.
7
Place the new battery in the base of the unit, in the same orientation as the original battery.
8
Place the strap over the battery.
9
Using a #2 Phillips screwdriver, install the screws to attach the strap to the base of the compartment.
10
Connect the power wires to the battery terminals, ensuring the wires are connected to the correct terminals:
►
Red (+) wire connected to the red (+) battery terminal
►
Black (-) wire connected to the black (-) battery terminal
11
Reinstall the base cover and secure with 6 screws using a #2 Phillips screwdriver.
10
Connect AC power cord to the power receptacle. Switch the AC ON/OFF switch
ON
. Switch the i.C
3
Monitoring System / Access Control backup battery ON/OFF switch
ON
.
9.4
Check Probe Bottle
Check solution:
1
Visually inspect the contents of the primary probe bottle.
►
Probe(s) should be immersed at least 2” (50 mm) in solution
►
If solution is low, clean and refill the probe bottle
Inspect and Replace Probe Bottle
NOTE
►
A kit that includes a probe bottle and glycerin is available from Haemonetics
Corporation BloodTrack
Customer Support.
Order Service Kit Part number 400922-1.
►
Grill must be removed to gain access to the probe and probe bottle.
►
Temperature probes are fragile; handle with care.
Required supplies:
►
4 oz. probe bottle
►
4 oz. (120 mL) of product simulation solution per bottle
►
10:1 ratio of water to glycerin
Inspect probe bottle:
1
Visually inspect the probe bottle for cleanliness, discoloration, and damage.
►
If probe bottle is damaged, replace the probe bottle and solution
►
If probe bottle is dirty, clean the bottle and refill the solution as outlined in
Section III, Item
9 .6 .7
Replace probe bottle:
1
Remove probe(s) from bottle.
2
Remove bottle from bracket.
3
Discard the bottle and solution.
4
Fill new bottle with 4 oz. (120 mL) of product simulation solution.
5
Cap bottle tightly to minimize evaporation.
6
Place bottle in bracket.
7
Replace probe(s), immersing at least 2” (50 mm) in solution.