10 .3
Maintaining Tray Assembly
10 .3 .1 Remove and Install a Tray and Cover
Please refer to the BloodTrack Courier
®
User Guide for details on preparing the HaemoBank™ refrigerator for service.
Remove a tray:
1
Open the exterior door, unlock, and rotate the Bypass Release handle counterclockwise 90 degrees.
2
Pull the tray out until it stops.
Tray fully extended
3
With one hand on each side of the tray, near the cover (at the rear of the tray), gently spread the top edges of
the tray outward.
4
While holding the top edges of the tray outward, use your thumbs to press the forward edge of the tray cover
upward.
►
The tabs on the outer edges of the tray cover will disengage from the corresponding slots on the top
edges the tray
►
The front of the tray cover will lift from the tray
5
Lift the tray cover upward.
►
The rear of the tray cover will lift from the tray
6
Remove the tray cover from the tray.
7
While keeping the tray level, lift the tray upward in the compartment location.
8
Pull the tray out of the compartment location.
Install a tray:
NOTE
The tray must be partially installed in the compartment location before the cover can be installed.
1
While keeping the tray level, slide the tray partially into the compartment location until the groove on the
bottom of the tray crosses the tray bumper.
2
Set the tray down in the compartment location.
3
Insert the tray cover into the compartment location, on top of the tray.
Insert tray cover
Posts oriented upward; tabs engaged in slots
►
The posts on the tray cover must be oriented upward
►
The tabs on the rear of the tray cover should engage the corresponding slots on the back of the tray