On-Board FastRegen Control Module Programming and Operation Instructions
P/N 8040406
C-1
C
HELIX TECHNOLOGY CORPORATION
-
TI
CRYOGENICS
Appendix C - RS-232 Regeneration
Responses
Introduction
Appendix C describes the On-Board Cryopump REGEN responses which
are requested by host computer O status command. The regeneration
responses inform the host computer of the regeneration cycle status of the
On-Board Cryopump.
NOTE: In some cases there are several responses which mean the same
thing.
Regeneration Response Description
Refer to Table C-1 for a complete description of the regeneration
responses.
NOTE: Regeneration responses identified with an asterisk (*) apply to
FastRegen responses. Regeneration responses not identified with an
asterisk apply to Full regeneration responses.
Responses to the O command are A - z (equal to numeric codes 0 - 57 and
58 - N). The numeric return information is preceeded by the plus sign (+).