On-Board FastRegen Control Module Programming and Operation Instructions
P/N 8040406
3-51
C
HELIX TECHNOLOGY CORPORATION
-
TI
CRYOGENICS
CONTROL Function Operation
Use the following procedures to operate the CONTROL parameters of the
On-Board FastRegen Control Module.
Cryopump ON/OFF and Second Stage Temperature
1. Press the CONTROL key on the keypad/display. The ON/OFF
status and the second stage temperature of the cryopump are
displayed.
NOTE: Refer to Appendix A and contact the local CTI-CRYOGENICS
customer support center if the letters SHO or OPN are displayed.
2. Press 1 to turn the cryopump ON, or press 0 to turn the cryopump
OFF.
3. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.
Aux TC ON/OFF and TC Pressure
1. Press the CONTROL key on the keypad/display. The ON/OFF
status and the second stage temperature of the cryopump are
displayed.
2. Press the Next key. The Aux TC pressure is displayed in microns.
3. Press 1 to turn the Aux TC ON, or press 0 to turn the Aux TC OFF.
4. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.
Cryo TC ON/OFF and TC Pressure
1. Press the CONTROL key on the keypad/display. The ON/OFF
status and the second stage temperature of the cryopump are
displayed.
2. Press the Next key until the Cryo TC pressure is displayed in
microns.
3. Press 1 to turn the CRYO TC ON, or press 0 to turn the Cryo TC
OFF.
4. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.