Programming and Operation
3-40
P/N 8040406
C
HELIX TECHNOLOGY CORPORATION
-
TI
CRYOGENICS
SERVICE Function Operation
Use the following procedures to operate the SERVICE function parameters
of the On-Board FastRegen Control Module.
Serial Number and Software Version
1. Press the SERVICE key on the keypad/display. The serial
number and software version information are displayed.
2. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.
Net Identification
1. Press the SERVICE key on the keypad/display. The serial
number and software version information are displayed.
2. Press the Next key. The Net ID number is displayed.
3. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.
Elapsed Pump Time
1. Press the SERVICE key on the keypad/display. The serial
number and software version information are displayed.
2. Press the Next key until the Pump Hours are displayed.
3. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.
Time Since Last Full Regeneration
1. Press the SERVICE key on the keypad/display. The serial
number and software version information are displayed.
2. Press the Next key until the Full RGN Hours are displayed.
3. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.