Programming and Operation
3-20
P/N 8040406
C
HELIX TECHNOLOGY CORPORATION
-
TI
CRYOGENICS
REGEN Function Programming
Use the following procedures to program individual REGEN function
parameters of the On-Board FastRegen Control Module.
Initiate REGEN
1. Press the REGEN key on the keypad/display. The current
REGEN status is displayed.
2. Press 1 to initiate a regeneration cycle.
3. Press 2 to confirm the start of a Full regeneration cycle.
4. Press 3 to confirm the start of a Fast regeneration cycle.
Abort REGEN
1. Press the REGEN key on the keypad/display. The current
REGEN status is displayed.
2. Press 0 to abort a regeneration cycle.
3. Press 2 to confirm the end of a regeneration cycle.
Delay Start
1. Press the REGEN key on the keypad/display. The current
REGEN status is displayed.
2. Press the Next key. The current delay start time value is displayed
in hours.
3. Press the Enter key. The current value is underlined and ready to
accept a new value.
4. Press the numeral keys for the desired delay time. For example: 1,
5 for 1.5, or 1, 5, and 0 for 15.0. The range of values is 0 - 999.9
hours.
NOTE: The Clear key may be pressed if incorrect data has been entered.
The On-Board software allows you to enter the new value.
5. Press the Enter key. Delay start programming is complete.
6. Press the MONITOR key. The On-Board microprocessor will
monitor system operation.