Programming and Operation
3-26
P/N 8040406
C
HELIX TECHNOLOGY CORPORATION
-
TI
CRYOGENICS
NOTE: If the cryopump temperature stays lower than the setpoint, the
cryopump is restarted. The cryopump has 30 minutes for the second stage
temperature to cool below 17K. If this temperature is not achieved, the
display will read POWER FAILURE - CHECK CRYO TEMP.
NOTE: If the cryopump is in regeneration and is cooling, the cooling will
continue if the thermocouple pressure is below 100µ. If the cryopump is in
any other phase of regeneration, or if the thermocouple pressure is above
100µ, regeneration will start over again. If cooldown continues, the
display will read POWER FAILURE - COOLDOWN.
Performing a Fast or Full Regeneration Cycle
Make sure you program the REGEN function parameters to your desired
values before attempting to perform a Fast or Full regeneration cycle.
Fast Regeneration Cycle
1. Press the REGEN key on the keypad/display. The current
REGEN status is displayed.
2. Press 1 to initiate a regeneration cycle.
3. Press 3 to confirm the start of a Fast regeneration cycle. A Fast
regeneration cycle has been initiated.
The On-Board FastRegen Control Module performs a Fast regeneration
cycle according to the REGEN parameters you have programmed. Refer to
Figure 3-10 for more information.
Full Regeneration Cycle
1. Press the REGEN key on the keypad/display. The current
REGEN status is displayed.
2. Press 1 to initiate a regeneration cycle.
3. Press 2 to confirm the start of a Full regeneration cycle. A Full
regeneration cycle has been initiated.
The On-Board FastRegen Control Module performs a Full regeneration
cycle according to the REGEN parameters you have programmed. Refer to
Figure 3-11 for more information.