5-36
95582716
OPERATION AND APPLICATION
KT19 II
20/06/19e
HEITRONICS Infrarot Messtechnik GmbH
5.2.14. Change analog output
The analog output of the instrument can be set to any temperature range or the corresponding radient
within the instrument measuring range. The smallest adjustable temperature span is 50 K.
The current temperature/radient value is assigned to the analog output by default. Alternatively minimum or
maximum value memory may be assigned to the analog output.
The output signal may be set to 0 to 20 mA, 4 to 20 mA, 0 to 1 V or 0 to 10 V.
Switch temperature to radiation proportional
5.2.14.1.
In case of a temperature linear output the output signal may be determined as linear interpolation between
the high and low configured temperature.
The radiant linear signal is a relatively calculated signal, depending on the detected radiance. In general
the knowledge about spectral characteristics of the instrument is necessary for this.
Call up menu
Select menu item
Confirm
Select menu item
Confirm
Select menu item
Adopt selection
Code setting see 5.2.4
Press several times or keep
pressing to return to measure-
ment