![Heitronics KT19 II Operation Instructions Manual Download Page 57](http://html1.mh-extra.com/html/heitronics/kt19-ii/kt19-ii_operation-instructions-manual_2120400057.webp)
5-9
95582716
OPERATION AND APPLICATION
KT19 II
20/06/19e
HEITRONICS Infrarot Messtechnik GmbH
Transmission/Reflection Mode (Setting of reflectance)
5.2.6.2.
Setting of reflectance indicates to what extent the ambient temperature is reflected
by the measured target and becomes a part of the received reflection.
Select menu item Emiss/Trans/Refl and activate Transmit/Reflect Mode by button
“USE”.
Emissivity (
𝜀
) is calculated by reflection (
𝜌
) and transmission (
𝜏
) by formula
ε = 1 − ρ − τ
.
Call up menu
Select menu item
Confirm
Select menu item
Use selection
Display and setting of reflection
and transmission values (if
activated)
Code setting see 5.2.4
Select menu item
Change value
Select digit
Change value
Save value
Further options (see
Option Intern/Extern/Manual)
Press several times or keep
pressing to return to measure-
ment