4.4.2
Aiming via focus laser
The illumination of the measured target diameter via the focus laser is switched on/off
the "keyboard".
At works the focus laser is set for the mounted lens. This means that here will be a reti-
cule with circle at the minimum target diameter for the mounted objective indicating the
target size and position. In case other objectives are attached to the Infrared Radiation
Thermometer, the focus laser is no longer adjusted, i.e. neither the position nor the size
of the target diameter is indicated correctly.
4.4.3
Aiming via pilot laser
ATTENTION
Please observe the SAFETY INSTRUCTIONS.
A pilot laser can be integrated into the Infrared Radiation Thermometer KT 19 II in case
a lens is used which is transparent to visible light. It marks the centre of the target di-
ameter for the objective which has been mounted at works. It is not possible to show the
minimum target diameter with this option.
4.4.4
Controlling anti-glare filter in optical view finder
Setting to continuous operation (IN) or automatic control by means of KT19 II.
Setting
:
FILter IN
↵
AUTo
4-6