5-33
95582716
OPERATION AND APPLICATION
KT19 II
20/06/19e
HEITRONICS Infrarot Messtechnik GmbH
Continuous measurement of ambient temperature (Option: External)
5.2.13.1.
For continuous measurement of the ambient temperature a 0 to 10 V signal must be fed in at terminal E
and B of the 12-pin plug connector corresponding to the configured ambient temperature.
Then the functional assignment of the input
is set to “Ambi. Temp” (ambient temperature) followed by the
configuration of the temperature span in menu “Emiss/Trans/Refl”.
Call up menu
Select menu item
Confirm
Select menu item
Use selection
Display and setting of ambient
temperature range
Code setting see 5.2.4
Change range
Select menu item
Change value
Select digit
Change value
Save value
Press several times or keep
pressing to return to measure-
ment
Same procedure valid for
High
For transmitting the measured value of a contact thermometer to a 0 to 10 V signal a recommended mod-
ule is necessary
See Options and Accessories
.