OPERATION AND APPLICATION
KT19 II
20/06/19e
HEITRONICS Infrarot Messtechnik GmbH
5.2.7.
Show and change response time
The response time of the instrument serves adjusting the reaction rate of the temperature indication and
corresponds to the time of integrating the measuring signal. The response time affects the noise figure and
thereby the associated temperature resolution. Selection of response time should reach a compromise
between time resolution and temperature resolution.
The response time may be selected between 0.05 sec and 600 sec. Temperature resolution is depending
on the spectral response. For more information see
chapter “Technical Specifications”.
Call up menu
Select menu item
Confirm
Change value
Code setting see 5.2.4
Select value
Confirm
Press several times or keep
pressing to return to measure-
ment