61
Troubleshooting
Alarm messages
Possible cause / remedy
Shock detected
Overload: Check the system configuration, restart the
system/process, contact the technical service if necessary.
Tilt F/B
and acoustic signal
S-box tilt in longitudinal direction too high: Check the
contents of the S-box, restart the system/process, contact
the technical service if necessary.
Tilt L/R
and acoustic signal
S-box tilt in transverse direction too high: Check the
contents of the S-box, restart the system/process, contact
the technical service if necessary.
Temperature out
of bounds
Temperature warning limit exceeded: Check the system
parameters/configuration, restart the system/process,
contact the technical service if necessary.
Co2 out of
bounds
CO
2
warning limit exceeded: Check the system parameters/
configuration, restart the system/process, contact the tech
-
nical service if necessary.
Humidity out of
bounds
Humidity warning limit exceeded: Check the system param-
eters/configuration, restart the system/process, contact the
technical service if necessary.
Battery
critical
Battery state of charge alarm limit exceeded: To charge the
battery, insert the S-box in the S-rack immediately.
Temperature not
reached in time
Target temperature not reached within the defined time:
Check the system parameters/configuration, restart the
system/process, contact the technical service if necessary.
CO2 not reached
in time
Set CO
2 concentration not reached within the defined time:
Check the system parameters/configuration, restart the
system/process, contact the technical service if necessary.
Humidity not
reached in time
Set humidity not reached within the defined time: Check
the system parameters/configuration, restart the system/
process, contact the technical service if necessary.
Out of memory
for events
Event memory overflow
: Save the event data externally and
empty the event memory.
Out of memory
for user-events
User data memory overflow
: Save the user data externally
and empty the
user data memory.
Out of memory
for series data
Measured data memory overflow
: Save the measurement
data externally and empty the
measured data memory.
Failure writing
event
Error on saving event data: Back up the measured data
externally and
empty the measured data memory, contact
the technical service if necessary
.
Failure writing
user-event
Error on saving user data: Back up the user data externally
and
empty the user data memory, contact the technical
service if necessary
.