Hybrimune Hybridoma Production System
Publication 015-1010191 Rev 4.0 • www.btxonline.com
System Test
The following is a basic test of the waveform generator system
using the preprogrammed protocol
HYBRIMUNE Test.pro
located
in the default \ protocol folder. For this initial test, the Chamber
Cable and Chamber should be fully connected, but it is not
necessary to fill the chamber.
First:
• Plug in the Line/Mains power cord for the
waveform generator.
• Turn the Line/Mains Power rocker switch at the lower
left front of the front panel to the ON (I) position
What should happen…
The rocker switch, “Power”, “Ready”, and
“Chamber Connected” LED’s should illuminate.
Troubleshooting tips:
If the rocker switch fails to illuminate, then return it to
the OFF position (O), and:
• Verify that the Line/Mains cord is properly
plugged into the receptacle on the back panel
and into the source of electric power.
• It may be necessary to check the fuse. See Maintenance
section for instructions, proper replacement of the fuse,
and the correct fuse type.
• If only the rocker switch and “Power” LED are
illuminated, then connect the Chamber Cable
and verify that the “Ready” and “Chamber
Connected” LEDs turn on.
Then:
Start the Waveform Generator Interface software.
What should happen…
The screen should appear like that as shown in
Figure 5. The CommLink window should show
OK
.
This display indicates that the computer is
communicating with the internal microprocessor of the
waveform generator. Additionally, the Electrofusion
Chamber window should display
OK
, the High Voltage
window should display
OFF
, and the Status window
should say
Ready
. The Power Supply Voltmeter window
will display 0 volts.
Troubleshooting tip:
If
ERROR
appears in the CommLink window, then turn
off the Waveform Generator and check the serial cable
connection to the waveform generator and computer
(or USB-to-Serial adapter, if applicable). Once
communication is established, the test may proceed.
Then:
• Click on the file folder icon on the left side of the screen
• Open the protocol file called
HYBRIMUNE Test.pro.
• Click Turn HV
ON
Figure 5: User Interface Application Opening Screen
Hybrimune
™
System Components & Set-Up