Chapter 5 – Configuration
Hardy H6800 Series Application Controllers User Guide
74
5.6 Calibrating the System
The HI6800 Series Application Controller must be calibrated upon installation. For more
information about calibration and other system setup parameters, see section 5.3.
Note:
Retail scales, used in legal or trade applications, must be verified and
sealed by an approved agency per the metrological directives. Please contact
Hardy Field Service with any questions.
5.6.1 Performing C2 Calibration
Path:
Home screen > Operation Menu > Calibration > C2 Calibration
C2 calibration calibrates a scale system electronically without using certified test weights. It uses
up to eight load sensors, a junction box, interconnect cable, and an instrument with C2
capabilities, such as the HI6800 Series Application Controller. Digital information within an HI C2-
certified load sensor details its unique performance characteristics. The HI6800 Series Application
Controller reads the performance characteristics of each load sensor and detects the number
of load sensors in the system.
Parameter
Description
Tare
The value you enter allows you to avoid pushing the Tare button each time you weigh something.
Range: .000001 - 999999
Default: 0.0
Zero
Weight unit limit from zero that the application controller accepts as gross zero during the zeroing function
(when you push the Zero button).
Range: .000001 – 999999
Default: 10.0
Mode
Weight source to use as the set point input.
Range: Net, Gross
Default: Gross
Calibration Date
Date when the calibration is performed.
Num Device
Read-only screen that shows the number of devices participating in the calibration.