Application measurement
EN
Copyright © Qingdao Hantek Electronic Co., LTD
HBT3000 series user manual
43
Internal loading setting steps: Set the loading path to internal, click on
Inter Load
,
and in the pop-up menu, there are a total of 6 states to choose from, ranging
from 0 to 5. Click the corresponding soft key to load it into the corresponding
startup settings. If there are saved startup settings in the current state, the screen
will display "StatX.bin Load Success!"; If there are no saved startup settings in the
current state, the screen will display "Load Failed StatX.bin Does Not Exist!".
External loading setting steps: Insert the USB flash drive into the USB HOST
interface on the front panel, set the loading path to external, click on
Exter Load
,
use up/down keys to select the. csv file, and click
[OK]
to load the file.
6.9.3
Boot Set
The startup settings include: Default, Last Off, 0, 1, 2, 3, 4, and 5. The default
selection for the instrument is default.
Press the
[LOAD] ->
Boot Set
soft key, select the desired configuration type, and
complete the power on settings.
6.10
0 ADJ
To eliminate errors caused by the bias voltage of the instrument itself or the
measurement environment, perform zero adjustment before measurement.
Before conducting the test, the instrument needs to be zeroed first. Before
zeroing, short-circuit the test line using the correct method, and then press the
[0ADJ] button to perform zeroing. The interface will display "Zero adjustment
Failed/Success" and disappear after 2 seconds.
Note
:
⚫
Perform zero adjustment on all ranges.
⚫
Even if the power is cut off, the compensation value after zero adjustment will
continue to be maintained.
⚫
Zero adjustment can be performed on the 0ADJ terminal of EXTI/O.
⚫
After successful zeroing, continue zeroing. If zeroing fails, retain the previous
zeroing result.
6.11
Local key
Click
[Local]
, the lock screen icon on the screen will light up, and clicking any
button other than this button is an invalid operation. Click the button again, and
the lock screen icon on the screen will turn gray.
In remote control mode, click
[Local]
to switch remote control to local control.