Application measurement
EN
HBT3000 series user manual
Copyright © Qingdao Hantek Electronic Co., LTD
28
function is turned off, there is a possibility of negative measurement values, which
can affect the judgment of comparison results.
Click on
[MEA] ->
Absolute
in sequence to set the absolute value, as shown in the
figure. Press the corresponding soft key again to complete the function of setting
absolute values.
Figure 6-6 Set absolute value function
6.2
Comparator function
The comparator function refers to the function of comparing pre-set critical values
with measured values, determining whether the measured values meet the
judgment criteria, and displaying and outputting them. The comparison methods
for thresholds include two methods: setting upper and lower limits, and setting
baseline values and percentages. As a comparator result, in addition to displaying
Hi, In, Lo pages and sounding the buzzer, it can also be output through the EXT
I/O terminal.
There are two comparison modes: Auto and Manual, with the default being Auto
mode.
Automatic mode: After setting to automatic mode and enabling the comparator
function, always output the comparison results of the comparator.
Manual mode: After setting to manual mode and enabling the comparator
function, the comparison result is only output when the MANU input of EXT I/O is
on.
Note:
When the comparator is turned on, automatic range cannot be used.
6.2.1
Comparator function settings
Click the
[Comp]
button to jump to the comparator settings page, as shown in
the figure.