Application measurement
EN
Copyright © Qingdao Hantek Electronic Co., LTD
HBT3000 series user manual
37
Figure 6-17 System settings
6.5
Calibration
Click on
[Utility] ->
Cal
in sequence, enter the password, and you can set [Zero],
[Calib Vol], and [Calib Res].
The specific steps are as follows
:
Click
Cal
to pop up the numeric keypad for the password.
After entering the correct password, enter the calibration submenu; If the
password is incorrect, a prompt message "Password Error" will pop up on the
interface.
Zero
Details reference
Appendix C: Performing Zero Adjustment
Voltage calibration setting steps
1
Connect the DC power supply to the input port of the internal resistance
tester and set the power output
2
Input actual voltage
Click on
Calib Vol
->
Real Vol
in sequence, and the page will display Voltage
Calibration (P). Use the left/right keys to move the cursor, click
[OK]
to enter the
actual voltage value, position the cursor to the key
, and then click
[OK]
to
save the value.