Application measurement
EN
HBT3000 series user manual
Copyright © Qingdao Hantek Electronic Co., LTD
38
Figure 6-18 Voltage Calibration
3
Set the next gear
Click
Next
to correct the next gear. Repeat steps 1 and 2.
When the page displays voltage calibration (N), the red and black probes should
be reversed and calibration should continue. After all gear calibration is
completed, the return menu in the bottom right corner switches to complete.
Click
Done
to complete the voltage calibration.
。
Steps for setting up resistance calibration
1
Connect the four wire resistor to the input port of the internal resistance
tester
2
Input actual resistance value
Click on
Calib Res
->
Real Res
in sequence, and the page will display Resistance
Calibration. Use the left/right keys to move the cursor, click
[OK]
to enter the
actual resistance value, position the cursor to the key
, and then click
[OK]
to
save the value.
3
Set the next gear
Click
Next
to correct the next gear. Repeat steps 1 and 2. After all gear
calibration is completed, the return menu in the bottom right corner switches to
complete. Click
Done
to complete the resistance calibration.
Note:
Before calibration, please preheat the machine for 30 minutes.
Before calibration, please confirm the rated values of the instrument and test
lines, otherwise it may cause damage to the instrument or personal injury
accidents.
6.6
IO
Communication settings include: LAN, Serial.