background image

25

Subject to change without notice

Prerequisite for this HF compensation is a square wave generator
with fast risetime (typically 4 ns), and low output impedance
(approx. 50 Ohm), providing 0.2 V at a frequency of approx. 1MHz.
The calibrator output of this instrument meets these requirements.

Connect the probe to the input previously used when 1 kHz
adjustment was made. Select 1 MHz output frequency. Operate
the oscilloscope as described under 1 kHz but select for 0.2 µs/
div time deflection coefficient setting.

Insert the probe tip into the output socket. A waveform will be
displayed on the CRT screen, with leading and trailing edges
clearly visible. For the HF-adjustment now to be performed, it will
be necessary to observe the rising edge as well as the upper left
corner of the pulse top. The location of the high frequency
compensation trimmer(s) can also be found in the probe
information sheet. These R-trimmer(s) have to be adjusted such
that the beginning of the pulse is as straight as possible. Overshoot
or excessive rounding is unacceptable. The adjustment is relatively
easy if only one adjusting point is present. In case of several
adjusting points the adjustment is slightly more difficult, but
causes a better result. The rising edge should be as steep as
possible, with a pulse top remaining as straight and horizontal as
possible.

After completion of the HF adjustment, the signal amplitude
displayed on the CRT screen should have the same value as
during the 1 kHz adjustment.

Probes other than those mentioned above, normally have a larger
tip diameter and may not fit into the calibrator output. Whilst it is
not difficult for an experienced operator to build a suitable
adapter, it should be pointed out that most of these probes have
a slower rise time with the effect that the total bandwidth of
scope together with probe may fall far below that of the
oscilloscope. Furthermore, the HF adjustment feature is nearly
always missing so that waveform distortion can not be entirely
excluded. The adjustment sequence must be followed in the
order described, i.e. first at 1 kHz, then at 1 MHz.

Prerequisites for precise and easy probe adjustments, as well as
checks of deflection coefficients, are straight horizontal pulse
tops, calibrated pulse amplitude, and zero-potential at the pulse
base. Frequency and duty cycle are relatively uncritical. For
interpretation of transient response, fast pulse rise times and low
impedance generator outputs are of particular importance.

Providing these essential features, as well as selectable output
frequencies, the calibrator of the instrument can, under certain
conditions, replace expensive square wave generators when
testing or compensating wideband attenuators or amplifiers. In
such a case, the input to an appropriate circuit will be connected
to the

 

calibrator output via a suitable probe.

The voltage provided by the probe to a high impedance input
(1 MOhm II 15 - 30 pF) will correspond to the division ratio of the
probe used (10:1 = 20 mV

pp

 output). Suitable probes are

 HZ51

,

52 

and

 54

.

Operating modes
of the Y amplifiers in Yt mode

The most important controls regarding the operating modes of
the Y amplifiers are the pushbuttons:

 CH I [15]

,

 DUAL [16] 

and

CH II [19]

. Their functions are described in the section ”Controls

and Readout”.

In most cases oscilloscopes are used to display signals in Yt
mode. Then the signal amplitude deflects the beam in vertical
direction while the time base causes an X deflection (from left to
right) at the same time. Thereafter the beam becomes blanked
and fly back occurs.

The following Yt operation modes are available:
1

st

Single channel operation of channel I (Mono CH I).

2

nd

Single channel operation of channel II (Mono CH II).

3

rd

Two channel operation of channel I and channel II (DUAL).

4

th

Two channel operation of channel I and channel II displaying
the algebraic result as the sum or difference  (“add”).

The way the channel switching is determined in

 DUAL 

mode

depends on the time base setting and is described in the section
”Controls and Readout”.

In

 ADD 

mode the signals of both channels are algebraically added

and displayed as one signal. Whether the resulting display shows
the sum  or difference is dependent on the phase relationship or
the polarity of the signals and on the invert function.

In

 ADD 

mode the following combinations are possible for

In phase input voltages:

    Channel II invert function inactive = sum.
    Channel II invert function active = difference.

Antiphase input voltages:

    Channel II invert function inactive = difference.
    Channel II invert function active = sum.

In the

 ADD 

mode the vertical display position is dependent upon

the

 Y 

position

 

setting of both channels. The same Y deflection

coefficient is normally used for both channels with algebraic
addition.

Please note, that the Y-position settings are also added but
are not affected by the invert function.

Differential measurement techniques allow direct measurement
of the voltage drop across floating components (both ends above
ground). Two identical probes should be used for both Y inputs.
In order to avoid ground loops, use a separate ground connection
and do not use the probe ground leads or cable shields.

X-Y Operation

The important control for this mode is the pushbutton labelled

DUAL

 and 

MENU [16]

.

In

 XY 

mode the time base is deactivated. The signal applied to the

input of channel I  front panel marking INPUT CHI (X)  causes the
X deflection. The input related controls (

AC/DC/GND 

pushbutton

and the

 VOLTS/DIV 

knob) consequently affect the X deflection.

For X position alteration, the

 X-POS.-

control knob must be used,

as the

 Y-POS./CURS.I 

control is automatically deactivated. The

input deflection coefficient ranges are the same for both channels,
because the

 X x10 

magnifier is inactive in

 XY 

mode.

Operating modes oft the Y amplifiers in Yt mode

Summary of Contents for HM504-2

Page 1: ...furbished and purchasing new Financing options such as Financial Rental and Leasing are also available on application TMG will assist if you are unsure whether this model will suit your requirements C...

Page 2: ......

Page 3: ...O s c i l l o s c o p e H M 5 0 4 2 Manual English...

Page 4: ...25 Phase comparison with Lissajous figures 26 Phase difference measurement in DUAL mode Yt 26 Phase difference measurement in DUAL mode 26 Measurement of an amplitude modulation 26 Triggering and tim...

Page 5: ...trol cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the in terior of the oscilloscope is screened by the cabinet direct radiation can...

Page 6: ...signal level Time Base delay provide high X Magnification of any portion of the signal 100 MHz 4 Digit Frequency Counter Cursor and Automatic Measurement Save Recall Memories for Instrument Settings R...

Page 7: ...ontrols Autoset automatic signal related parameter settings Save and Recall 9 instrument parameter settings Readout display of menu parameters cursors and results Autom Measurement Freq Period Vdc Vpp...

Page 8: ...the handle can be set into diffe rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked STOP Attention When changing t...

Page 9: ...nrequires the horizontal or inclined position STOP Do not obstruct the ventilation holes Specifications are valid after a 20 minute warm up period between 15 and 30 degr C Specifications without toler...

Page 10: ...pprox 10 and the real voltage value is 11 higher The gain reduction error can not be defined exactly as the 3 dB bandwidth of the Y amplifiers differs between 50 MHz and 55 MHz When examining square o...

Page 11: ...pe However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 2...

Page 12: ...nt lowerinputsensitivity must be chosen until the vertical signal height is only 3 8 div With a signal amplitude greater than 160 Vpp and the deflection coefficient VOLTS DIV in calibrated condition a...

Page 13: ...eristic resistance of the cable as a rule 50 Ohm Especially when transmitting square and pulse signals a resistor equal to the characteristic impedance of the cable must also be connected across the c...

Page 14: ...d by reducing the READOUT intensity RO The INTENS FOCUS control knob adjusts the READOUT intensity Turning this knob clockwise increases and counter clockwisedecreasestheintensity Onlytheminimumrequir...

Page 15: ...Rot with Int trace rotation by use of INTENS FOCUS knob and allows you to compensate the influence of the Earth s magneticfieldonthetracedeflection Toavoidmisadjustment due to unavoidable deflection...

Page 16: ...selected This allows for a different slope setting for the triggered DELAY DTR time base mode 10 TR Trigger indicator LED The TR LED is lit in Yt mode if the triggering conditions are met for the firs...

Page 17: ...the readout e g Y1 deflection coefficient input coupling The symbolizes calibrated measuring conditions and is replaced by the symbol in uncalibrated conditions VAR The vernier variable function is de...

Page 18: ...e internal trigger signal originating from channel I or channel II or the external trigger signal can be chosen DUAL and Addition mode Briefly pressing opens the trigger source Pulldown menu with the...

Page 19: ...s the way the trigger signal is connected to the trigger unit AC DC content suppressed DC peak value detection inactive HF high pass filter cuts off frequencies below approx 50 kHz trigger point symbo...

Page 20: ...e trace start position which is present after switching over from sea to del This enables lower time deflection coefficient settings for signal expansion del In DELAY mode a trigger event does not sta...

Page 21: ...PUT CH II BNC socket This BNC socket is the signal input for channel II The outer ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of t...

Page 22: ...ifferent frequency responses of the trigger amplifier and Y amplifier cause a reduction of the measurement accuracy If relatively low frequency signals 20 Hz are present the measurement value continuo...

Page 23: ...io X measurement causes the display of two long and one short CURSOR lines and is enabled inYt time base mode only The unit to be displayed must be selected by briefly pressing the UNIT 35 pushbutton...

Page 24: ...es are displayed Briefly pressing SOURCE selects the channel and it s deflection coefficient for the measurement The CURSOR lines must be set to the signal according to the selected channel b DUAL mod...

Page 25: ...ubmenus Main Menue 50 Ohm For high impedance loads Oscilloscope approx 1 M Ohm Digital Voltmeter approx 10 MOhm the output voltage is either 0 2 Volt DC or 0 2 Vpp AC square wave Under First Time Oper...

Page 26: ...n generator provides a square wave signal with selectable frequencies and a very fast rise time 4 ns from the output socket below the CRT screen As the square wave signals are used for probe compensat...

Page 27: ...the calibrator output via a suitable probe The voltage provided by the probe to a high impedance input 1 MOhm II 15 30 pF will correspond to the division ratio of the probe used 10 1 20 mVpp output S...

Page 28: ...the phosphor at too high a brightness setting INTENS setting which causes either a lasting loss of brightness or in the extreme case complete destruction of the phosphor at this point Phase difference...

Page 29: ...a sine signal Except when external trigger is used the trigger threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger...

Page 30: ...formation regarding this item can be noted in the Data Sheet The coupling setting TRIG MODE 20 and indication are described under Controls and Readout As the automatic triggering does not work below 2...

Page 31: ...er threshold less than 0 5 div It is therefore particularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in a circuit In this trigger mode the slope di...

Page 32: ...gnals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use of the holdoff control simplifies t...

Page 33: ...osen in del and dTr mode is stored and automatically set after activating one of those modes If the stored time coefficient in del dTr mode was higher than the actual value in sea search mode the time...

Page 34: ...me base mode I X x10 magnifier switched off I Optimum X and Y position settings I Trace and readout visible If DC trigger coupling had been selected AC trigger coupling will not be chosen and the auto...

Page 35: ...w typical patterns displayed by the various components under test I Open circuit is indicated by a straight horizontal line I Short circuit is shown by a straight vertical line Testing Resistors If th...

Page 36: ...or ground for the same terminal is then absolutely necessary A connection inversion effects a rotation of the test pattern by 180 degrees about the centre point of the scope graticule Pay attention t...

Page 37: ...ation The oscilloscope is supplied with a serial interface for control purposes The interface connector 9 pole D SUB female is located on the rear of the instrument Via this bidirectional port the ins...

Page 38: ...RECALL INPUT CHI X AC DC GND max 400Vp 1M II 18pF INPUT CHI X max 400Vp 1M II 18pF TRIG EXT max 100Vp INP Z x1 x10 AC DC GND x1 x10 50 MHz ANALOG OSCILLOSCOPE HM504 2 Instruments CAT I TRACE ROT CT GL...

Page 39: ...37 Subject to change without notice Front Panel HM504 2...

Page 40: ......

Page 41: ...LEER...

Page 42: ...chwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes Spectrum Analyzer P...

Reviews: